Pipette Calibration and Cleaning


Why choose wikiHow?
When you see the green expert checkmark on a wikiHow article, you know that the article has received careful review by a qualified expert. If you are on a medical article, that means that an actual doctor, nurse or other medical professional from our medical review board reviewed and approved it. Similarly, veterinarians review our pet articles, lawyers review our legal articles, and other experts review articles based on their specific areas of expertise.

How to Do Pipette Calibration

Two Parts:

Pipettes are a type of lab equipment used to measure and transfer very small volumes of liquid. Accuracy and precision in pipette measurement are necessary as any discrepancy in volumes dispensed can affect the results of an experiment. To ensure accuracy, it is necessary to check pipette calibration every few months. The calibration process helps to check whether or not the equipment is dispensing the proper volumes so it can be fixed if it is not.

Steps

Checking the Calibration

  1. Gather the necessary materials.To check the calibration of a pipette, you will need the pipette, pipette tips, distilled water, a beaker, a thermometer, a balance, and weigh boats. The balance needs to be specific to micrograms to calibrate micropipettes with a maximum of 1 µL.
    • You won’t need more than 5 mLs of water. Fill the beaker with the water.
    • Make sure the pipette tips are the correct tips that properly fit the pipette.
  2. Measure the temperature of the distilled water.Place the thermometer in the water and leave it for at least one minute. If the red line on the thermometer is still moving, leave it for another minute. After a minute, record the temperature in a notebook. Remove the thermometer and dry it off when you are finished.
    • Knowing the temperature of the water is important for the calculations that will be done to check the calibration.
  3. Place the weigh boat on the balance and zero it out.Ideally, you will use a balance that has doors on it and is an isolated chamber. Place the weigh boat in the chamber and close the doors. If you don’t have doors, simply place the weigh boat on the balance. Press the “Zero” or “Tare” button and wait until the balance reads zero.
    • Zeroing the scale subtracts the weight of the plastic boat and allows you to measure just the weight of whatever you put into the boat.
  4. Prepare the pipette for calibration.Wipe down the pipette with ethanol to kill any contaminants before beginning and make sure nothing is clogging the tip end of the pipette. Place the appropriate tip on the end and set the volume that you want to test.
    • For calibration, test both the smallest volume and the largest volume the pipette can dispense.
  5. Pre-rinse the tip before calibration.Press the button to the first stop and stick the tip into the distilled water so that about 2 mm is immersed in the liquid. Release the button to aspirate some liquid and then dispense the liquid again by pushing the button down. Repeat this three times to pre-rinse the tip.
    • Push the button to the second stop to dispense any liquid that remains in the tip and remove the pipette from the water.
  6. Aspirate the calibration volume.With the tip out of the liquid, push the button down to the first stop. Place about 2 mm of the tip into the distilled water and release the button to aspirate the liquid into the tip. Wait about 1 second before removing the tip from the water.
    • Make sure the tip is completely immersed during the whole aspiration process. You don’t want any bubbles in the tip or the results will be wrong.
  7. Dispense the liquid into the weigh boat on the scale.Place the tip against the bottom of the weigh boat and push the button down to the first stop. Move to another spot slightly away from the water and push the button to the second stop. Keeping the button pressed, lift the tip away from the weigh boat.
    • Keep the tip on the pipette as you will use it again to weigh out more calibration tests. MAke sure the tip does not touch anything.
  8. Record the weight on the balance.Close the door if you are using a chamber balance. Wait until the numbers stop changing. Record this number in your notebook.
    • It’s important to wait until the number has stabilized before recording it. Your readings will be incorrect if you don’t wait.
  9. Repeat this process to take at least 10 readings.Zero the scale, pre-rinse the tip, aspirate the same volume, dispense the volume, and then record the weight. You want to have multiple recordings of the same volume and then you can average them all together.
    • You can repeat this same process with different volumes as long as you take multiple readings per volume.

Calculating the Results

  1. Define the formula for calculated volume.The formula for calculating the volume dispensed by the pipette is V = w * Z where w is the weight of the water, Z is the conversion factor based on the density of the water, and V is the calculated volume of how much water was dispensed.
    • You can find the Z variable by looking up the density of water using the temperature recorded at the beginning of the experiment.
    • For example: If the temperature of the water was 23°C, then you will use a Z value of 1.0035 µg/mg.
  2. Average the replicates together.You should have weighed the volume of water dispensed by the pipette at least 10 times. In order to average all of these values together, add them all up and divide by 10. If you did more or fewer trials, add up each trial and divide by the total number of trials.
    • For example: The ten weights you recorded for a pipette volume of 10µL are as follows: 9.89, 10.01, 10.02, 9.99, 9.95, 10.04, 9.96, 10.01, 9.99, and 9.98.
    • The average is: (9.89 + 10.01 + 10.02 + 9.99 + 9.95+ 10.04 + 9.96 + 10.01 + 9.99 + 9.98)/10 = 99.84/10 = 9.984
  3. Plug the variables into the equation and solve.Once you have determined the proper numbers to use for each variable, you can plug them into the equation and solve for the calculated volume. To solve, simply multiply the average weight of all the trials by the Z value.
    • For example: V = w * Z = 9.984 * 1.0035 = 10.019
  4. Calculate the accuracy of the pipette.To calculate the accuracy you can use the equation A = 100 x Vavg/V0, where A is the accuracy of the pipette, Vavgis the average calculated volume and V0is the value you set the pipette to dispense. Accuracy should be between 99-101%.
    • If the pipette is properly calibrated, the calculated value should be very close to the actual value you set on the pipette.
    • For example: A = 100 x Vavg/V0= 100 x 10.019/10 = 100 x 1.0019 = 100.19%
    • This particular pipette is properly calibrated.
  5. Send the pipette for calibration, if necessary.If your pipette does not pass the calibration test, stop using it for experiments immediately. Pipettes are very delicate and expensive pieces of lab equipment. You cannot fix the calibration yourself, so you will need to send them in for proper servicing. Alternatively, some companies will come to your lab and calibrate your pipettes there.
    • Contact the company that makes your particular brand of pipette for servicing.

Community Q&A

Search
  • Question
    How do I calibrate the instrument for use at a temperature of 24 degrees Celsius?

    Ph.D. candidate in Health Sciences
    Meredith Juncker is a Ph.D. candidate in Health Sciences at Louisiana State University.
    Ph.D. candidate in Health Sciences
    Expert Answer
    If the distilled water you used was 24 degrees Celsius at atmospheric pressure, your Z value would be 1.0038 µg/mg. Multiply these two values together (w*Z) to obtain your V value, or volume of water dispersed. This V value you just calculated is your Vavg. Divide Vavg/Vo (the amount you're theoretically pipetting, in this case it would be 1mL) and multiply this number by 100. That will be your accuracy percentage. Any answer between 99-101% is properly calibrated.
    Thanks!
  • Question
    Does this process apply to fast release pipette pumps?

    Ph.D. candidate in Health Sciences
    Meredith Juncker is a Ph.D. candidate in Health Sciences at Louisiana State University.
    Ph.D. candidate in Health Sciences
    Expert Answer
    Fast-release pipette pumps are made for pipettes that are calibrated to the tip but retain the final drop. They are usually calibrated professionally by the company that manufactures the pump.
    Thanks!
  • Question
    What is the reference standard for pipette calibration?
    wikiHow Contributor
    Community Answer
    The reference standard is ISO8655. This is a verification of your pipette readings, but, if you require a valid certificate, the calibration should be done by an accredited calibration laboratory IEC/ISO 17025.
    Thanks!
  • Question
    What are the factors to be considered for calculating the uncertainty of a pipette?
    wikiHow Contributor
    Community Answer
    You're confusing the human concept of uncertainty with the scientific concept. For a pipette, uncertainty isn't expressed in terms of conceptual sources of uncertainty. It's expressed in mathematical terms based on the standard deviation of a set of measurements.
    Thanks!
Unanswered Questions
  • Do the table Z values (g/mg) apply similar as when your weight units are in grams, not milligrams, and the pipette volume is in milliliters instead of micro-liters?
  • How to get uncertainty value for 100-1000 micropippete?
  • Where can I learn about repair and calibration of pipettes?
  • Short note of calibration of pipette?
Ask a Question
200 characters left
Include your email address to get a message when this question is answered.

Related wikiHows

Sources and Citations

Article Info

Categories:

In other languages:

Español: , Русский: , Português: , Italiano: , Bahasa Indonesia: , Français: , Deutsch: 

Thanks to all authors for creating a page that has been read 139,378 times.




Video: Calibration and Use of a Volumetric Pipet

How to Do Pipette Calibration
How to Do Pipette Calibration images

2019 year
2019 year - How to Do Pipette Calibration pictures

How to Do Pipette Calibration advise
How to Do Pipette Calibration forecasting photo

How to Do Pipette Calibration pics
How to Do Pipette Calibration photo

How to Do Pipette Calibration How to Do Pipette Calibration new images
How to Do Pipette Calibration new foto

photo How to Do Pipette Calibration
pictures How to Do Pipette Calibration

Watch How to Do Pipette Calibration video
Watch How to Do Pipette Calibration video

Discussion on this topic: How to Do Pipette Calibration, how-to-do-pipette-calibration/
Communication on this topic: How to Do Pipette Calibration, how-to-do-pipette-calibration/ , how-to-do-pipette-calibration/

Related News


The 5-Move Workout That’ll Help You Blast Cals Even When Your Kids Won’t Leave You Alone
Can this woman change the way you eat forever
How to Set Up a DLink WBR2310 Wireless Router
Resume for Someone Looking for a Second Job
How to Dispose of a Damaged American Flag
Kelly Hoppen sues News of the World over alleged phone hacking
Brown Sugar and Ginger Spice Cakes Recipe
Salvatore Ferragamo celebrates 80th anniversary
Duchess Kate Reveals What The Royal Family Tease Her About
23 Cool Finger Tattoo Ideas
The Heartbreaking Story Behind This Dads Postpartum Depression Foundation
Dolce Gabbana True Monica Spring 2013 Makeup Collection
How to Give Your Lonely Dog Some Company




Date: 03.12.2018, 23:13 / Views: 31172